Products

Tel: (+86) 531 58056101

Fax: (+86) 531 58756239

Mail: sales@metobt.com

Skype: sales@metobt.com

C64.605 Hydraulic Servo Universal Testing Equipment/Hydr


C64.605 Hydraulic Servo Universal Testing Equipment/Hydr

Load frame configuration: 6 column, servo control hydraulic system
Rated load: 600kN
Test space: double space (space stretching, small space compression)
Typical specimen: metal sheet, composite material, pipe and tube material, steel bar, welding parts, casting, concrete pavement

Technical parameters:

 

Specification model

C64.605

Rated load

kN

lbf

600

135000

 

Column configuration

Column number

6

Test space

Single/double

double

Actuator (piston) stroke

mm

in

200

7.87

Actuator (piston) speed

Mm/min

In/min

0.5 - 1400.0004-5.51

Beam velocity

mm/min

in/min

210

8.27

Column spacing (test space width)

mm

in

430

16.93

Maximum compression space

 

 

Standard length

mm

in

77030.31

Extended length

mm

in

112044.09

Maximum stretch space

 

 

Standard length

mm

in

75029.53

Extended length

mm

in

110043.31

Diameter of circular specimen

mm

in

10-40

0.39-1.57

Thickness of flat specimen

mm

in

2-30

0.07-1.18

Pressure plate

mm

in

150 x 150( square)

5.91 x 5.91( square)

Frame height

 

 

Standard length

mm

in

2390

94

Extended length

mm

in

2780109.45

Frame width

mm

in

1170

46.06

Frame depth

mm

in

800

31.50

Frame weight

 

 

Standard length

kg

Lb

3150

6945

Extended length

kg

Lb

3254

7174

 

Calibration standard

ISO 7500 0.5level or ASTM E4

ISO 7500 1level

load range

1%-100%The rated load

0.5%to1% The rated load

The displacement resolution

0.2μm

(0.00000787in)

Accuracy of displacement

The value for +/-1%

Accuracy of strain

ASTM E83 ISO 8513

Security and protection

Overload protection, limit, hydraulic overload protection,
overheating protection, voltage, overload protection and other

Force overload protection

10%

Data acquisition frequency

1000Hz

Control loop frequency

1000Hz

Environmental requirements

 

Working temperature range

5 to 40

41 to 104°F

Working humidity

5-85% non-condensing

Storage temperature range

-18 to 49

0-120°F

The largest storage humidity

90% non-condensing

Maximum altitude

2000 meter

Integrated operating platform (IOP) power supply requirements

400V 50Hz/480V 60Hz V AC

(3 phase)

IOP size

mm

in

1040 x 720 x 1900

38.58 x 28.35 x 74.8

IOP weight

kg

Lb

400

881.85

 

 

Main characteristics of C60 series system
1: high stiffness 6 column load frame configuration
2: reliable MTS servo control hydraulic system
3: high resolution, digital closed-loop controller
4: easy test setup and handheld
5: multi function, easy to operate the TestWork software, with rich and increasing test method standard library (ISO, ASTM, DIN, EN, BS, etc.)
6: a full set of accessories, alternative, fixture system and extensometer
7: compact and ergonomic integrated operating platform
8: standard double test space to reduce setup time
9: on space stretching
10: lower space compression
11: stepless loading
12: "rewind" hydraulic valve, in order to improve work efficiency
13: beam displacement limit protection, overload protection, over heat protection, voltage overload protection, etc.
14: optional EU standard independent safety protection cover
15: fully comply with the mechanical device 2006/42/EC, low voltage 2006/95/EC, EMC directive 2004/108/EC and GOST-R
16: durable test space protection
17: easy to maintain